TEST METHOD FOR HYDRAZINE IN INDUSTRIAL WATER (RANGE: 0-500 µg/L or say 0-0.5 mg/L N2H4)

BACKGROUND OF THE TEST:

Hydrazine having chemical formula as N2H4 has numerous applications such as oxygen scavenger to 

cater corrosion attack in the boilers circuits, for synthesis of various pesticides, in rockets fuels, etc. 

Nuclotrasis offers user friendly test method with ease of handling the relevant reagents and other 

equipments used in the analysis. For determination of hydrazine, this test relies on Lambert Bear’s 

Law of spectrometry where para-dimethylaminobenzaldehyde in acidic medium is allowed to react 

with hydrazine developing yellow color complex that is directly proportional to the hydrazine 

concentration in the solution. In this test, wavelength for maximum absorbance is 455 nm for 

spectrophotometers and 420 nm for colorimeters.

REAGENTS/EQUIPMENTS/APPARATUS:

 Provided 10 mL marked glass vial (Item Code: 02-01)

 Any Suitable Spectrophotometer or available softcopy of color comparative chart 

 Relevant Cuvette or cell of Photometer

 NUCLOTRASIS HYDRAZINE TEST SOLUTION (Item Code: 01-01)

SAMPLE COLLECTION & PREPARATION

 Make sure to immediately analyze the sample after collection of sample.

 Ensure to collect sample in a suitable and designated bottle for a specific sample.

 Thoroughly rinse the sample bottle with current sample and fill the sample by allowing 

slight overflow and immediately tighten the cap.

 Prevent agitation of the sample and exposure to air as the test is sensitive to the air.

TEST PROCEDURE:

1. Take sample to 10 mL mark into the provided vial (Item Code: 02-01) and call it as “A”.

2. Similarly take 10 mL deionized water in another cuvette or test tube for reagent blank and 

call it as “B”.

3. In the sample, add 20 drops (1 mL) NUCLOTRASIS HYDRAZINE TEST SOLUTION and mix 

gently.

4. Allow to stand for ten (10) minutes for full color development. 

5. Meanwhile, select Hydrazine specific channel/test in the spectrophotometer.

6. First run reagent blank “B” to achieve blanking.

7. Then run test sample “A” and note the Photometer display reading in µg/L or mg/L. In case 

photometer is not available, match the developed color with the provided soft copy color 

comparator of this HYDRAZINE test kit.

8. Multiply the reading with dilution factor 1.1.

REFERENCES: ASTM Manual of Industrial Water, D1385-78, 376 (1979)